calibration similar to one conducted during production process. To perform the accuracy correction, select corr from the CAL menu.

Perform correction according to required measurement parameters by pressing Action button, see table below:

Active mode of

Correction

 

measurement

Condition

parameter

(Std)

 

 

 

 

 

 

D, dE

Distance

Install unit at 1 meter1

from a flat target2

C

 

Put unit on a level plane

Azimuth

and aim it to the North

 

 

direction 3

E, dE

Elevation

Put unit on the

horizontal level

 

 

 

 

 

1 – Accuracy of correction will depend on accuracy of installation against the target. It must be exactly as specified. The numbers shown on the display reflect the correction.

2 – Target should not have high-reflective surface (white color is not considered high reflective in this case).

3 – The unit shall be well levelled with deviation from 0 not more than ±0.75 degrees. Horizontal direction shall be to

15

Page 19
Image 19
Newcon Optik LRM 220SI operation manual Std