SIGNAL INPUT, module ME

OVERVIEW

This module together with the "MP" input board, performs all the input signal conditioning (see Fig. 2). This module contains the trimmers and jumpers for the amplifier gain and the low level output (offset).

The signal input connections are made at Terminals 3, 4, 5 and 6.

Each model may use some of these terminals, see the Connections Section.

FIG.7

 

1

 

P1

2

P2

 

 

P1 : Output zero adjustment.

P2 : Gain amplifier adjustment.

B

A

C

Solder pads on solder side

OFFSET ADJUST

Solder pad 1 if closed Solder pad 2 if closed Jumper A if closed

:Adjust the low range level of the output (Offset positive coarse).

:Adjust the low range level of the output (Offset negative coarse).

:Adjust the low range level of the output (Offset negative fine).

AMPLIFIER GAIN

Jumper B if closed Jumper C if closed Jumpers B and C opened

:Gain at maximum level

:Gain at medium level.

:Gain at minimum level.

8