(OMWT-ANGOUT)

online (R) - indicates if the device is communicating (0 or 1) polltime (R/W) - how often to poll the device (in milliseconds) nodaladdress (R/W) – (TCP/IP Receiver). The node address starts at 1.

For TCP/IP Receivers, the node address corresponds to the node number for the TCP/IP connection. You can write to this parameter to determine which node to send the output packet. (write 0 and OMWT- SOFT-OPC will send to all nodes)

repeateraddress (R) – (Wireless mode) contains the location identifier of the OMWT-REPEAT Repeater. If the sensor did not pass through a repeater then the value is 0.

serialno (R) - registration number of the access/control reader updatecount (R/W) – value is incremented whenever the sensor is updated. output.value (R/W) – set variable to the desired Engineering value output.raw (R/W) – set the variable to the binary value (0 to 4095). output.nits (R/W) - units label assigned to the analog output. output.conv (R/W) - unit conversion parameters string.

(Cal1Raw,Cal1Engr,Cal2Raw,Cal2Engr)

out.status (R) – 1 – OMWT-SOFT-OPC is in the process of sending output packets based on tries and interval; 0 – OMWT-SOFT-OPC has finished sending the output packet.

out.interval (R/W) – time in milliseconds between OMWT-SOFT-OPC sending the output packet.

out.tries (R/W) – the number of times OMWT-SOFT-OPC will send the output packet. “-1” – never stops sending packets.

out.service (W) – When written to, OMWT-SOFT-OPC sends output packets to the output device for training.

Note: when either output.value or output.raw is written to, OMWT-SOFT-OPC will send out output packets based on outtries and outinterval. It first sets outservice to 1, and then to 0 when finished.

(R) – indicates that the tag is readable.

(W) – indicates that the tag is writeable only.

(R/W) – indicates that the tag is both readable and writeable. OPC Write operations will either change an I/O point at the device, or an operation will be performed.

OMWT-SOFT-OPC Server

OMWT-SOFT-OPC Server Tags N 43

Page 47
Image 47
Omega Engineering M-4066 manual Wireless Analog Output OutputAnalogn, Omwt-Angout

M-4066 specifications

Omega Engineering, a renowned name in measurement and control solutions, has developed a range of high-quality devices to meet the diverse needs of industries worldwide. Among these products, the Omega M-4066 stands out as a versatile and highly functional instrument designed for precise measurement and control applications.

The Omega M-4066 is a state-of-the-art digital thermometer, ideally suited for various industrial applications, including laboratory settings and process control environments. With its robust construction and user-friendly interface, this device is designed to facilitate accurate temperature readings in real-time.

One of the standout features of the M-4066 is its wide temperature measurement range. It can accurately measure temperatures from -200°C to 1372°C, making it suitable for a variety of applications, including cryogenics and high-temperature processes. This versatility allows users to apply it across multiple sectors, including manufacturing, pharmaceuticals, and food processing.

Equipped with advanced thermocouple technology, the M-4066 ensures rapid response times and high accuracy in temperature measurements. Its compatibility with different types of thermocouples, such as Type K, J, T, and E, enhances its versatility, allowing users to select the most appropriate sensors for their specific applications.

The M-4066 also features a large, easy-to-read LCD display that provides clear and precise temperature readouts. A range of additional functionalities, including maximum and minimum temperature readings, allows users to monitor fluctuations and maintain the desired environmental conditions effectively.

In terms of connectivity, the device offers options for data logging and communication with external systems. Users can interface the M-4066 with computers or control systems, enabling them to analyze data trends and implement necessary adjustments in real time.

The Omega M-4066 is built with durability in mind, featuring an IP65-rated enclosure that protects against dust and water ingress. This resilience is particularly beneficial for use in harsh industrial environments, ensuring longevity and reliable performance.

In summary, the Omega M-4066 is a powerful, flexible, and reliable digital thermometer that provides accurate temperature measurement across a broad range of applications. With its advanced features, diverse connectivity options, and robust design, it is an essential tool for professionals looking to enhance their measurement and control capabilities.