Check once a month whether the filter in the filter compartment is still white and clean. If it turns dirty, replace it. However this filter should be replaced at least once a year.
Procedure for replacing:
1. Slide the filter cover open to expose the filter compartment.
2. Remove the dirty filter with a small, pointed object.
3.Replace with a clean OMRON filter (order no.8403). A set of 5 spare filters is provided with the nebuliser.
4.Slide the filter cover closed.
Caution: Do not use cotton or any other material to replace the filter. Do not wash or clean the filter. Use only OMRON filters. Do not operate without a filter.
Possible problems and how to solve them
If the OMRON CompAIR CXpro does not nebulise, or only very weakly, check the following points:
●Make sure that the power supply plug is firmly fitted onto the wall socket.
●Make sure that the ends of the air tube are fitted tightly onto the compressor unit and onto the medication tank.
●Check whether the medication tank is empty. If so, and if necessary, refill the medication tank.
●Is the vaporiser head installed correctly in the medication tank? If not, install it and press it down into the jet nozzle.
●Check whether the air tube is twisted. If so, untwist.
●Check whether the jet nozzle is blocked. If so, unblock it by cleaning the medication tank. If not possible to unblock, replace the nebuliser kit.
●Check whether the air filter is clean. If not, replace it (refer to section 'Filter' in this manual).
Optional accessories/ parts:
Nebuliserkit JetAIRplus, (incl. airtube, mouthpiece, |
| |
inhalation/exhalationvalves), | order no. | A040001 |
Inhalation valve for JetAIRplus, | order no. | P040017 |
Exhalation valve for JetAIRplus, | order no. | P040018 |
Vaporiser head for JetAIRplus, | order no. | P040026 |
Airtube for JetAIRplus, | order no. | P040019 |
Flexible adult mask MAF, | order no. | A040002 |
Flexible child mask MPF, | order no. | A040003 |
Airfilter (set = 5 pcs), | order no. | 8403 |