Other problems

Error

Priority

Cause

Solution

code

 

 

 

 

 

Inflated the cuff to 300 mmHg or more

When inflating in "Auscultation Mode",

 

 

release the button when the pressure

 

 

during inflation in "Auscultation Mode".

 

 

reaches the desired value.

 

 

 

 

 

 

 

E3

Low

 

If this occurs during measurement, repeat

 

measurement.

 

 

 

 

 

Over-inflation occurs

If this occurs when not performing

 

 

 

measurement, contact OMRON

 

 

 

HEALTHCARE Customer Service.

 

 

 

 

E40

Medium

The battery is depleted.

Recharge the battery, or replace with a new

battery. (page 14)

 

 

 

 

 

 

 

E41

Medium

Battery failed to charge.

Try charging again. If continues to fail,

replace with a new battery. (page 14)

 

 

 

 

 

 

 

 

 

 

Replace the battery with a new battery. If the

E42

Medium

Battery voltage error

error continues, contact OMRON

 

 

 

HEALTHCARE Customer Service.

 

 

 

 

Disposal

Description

As there is a risk of environmental pollution, follow your applicable Federal, state and local legal regulations regarding disposal or recycling of this equipment and batteries.

The main constituents of each part are listed in the table below. As there is a risk of infection, do not recycle patient attachments such as cuffs, but dispose of them as instructed by your facility's procedures and applicable regulations.

Item

Parts

Material

 

 

 

Box

Cardboard

 

 

Package

Cushion

Cardboard

 

 

 

Bag

PE

 

 

Main unit and

Enclosure

ABS, PC , SR

 

 

accessories

Internal parts

General electronic components

 

 

 

Outer tube

PVC

 

 

Battery pack

Cell batteries

Nickel-hydrogen battery

 

 

 

 

 

 

 

 

Internal parts

General electronic components

 

Maintenance

 

 

 

 

29

Page 31
Image 31
Omron HBP-1300 specifications Disposal, Other problems, Description, Error Priority Cause Solution Code, Parts Material