Omron HBP-1300 Stethoscope, Abnormal measurement value, Measurement value is questionable

Models: HBP-1300

1 40
Download 40 pages 60.07 Kb
Page 29
Image 29

Abnormal measurement value

Cause / solution

The causes below are possible. Check the patient by palpation and then repeat measurement.

Body movement (chills or other trembling)

Arrhythmia.

Noise in the cuff

-A nearby person touched the patient.

-Cardiac massage was being performed.

The measurement value is questionable

Cause

 

Solution

 

 

 

Deflates quickly

 

Check for a loose cuff connection.

 

 

 

Stethoscope

Simultaneously perform measurement with a stethoscope.

 

Place the stethoscope and listen while viewing the pressure display of the

 

manometer.

Blood pressure may vary widely due to physiological effects.

The causes below are possible.

Emotional excitement or agitation

-Pain due to cuff wrapping

-White coat hypertension

Cuff size or wrapping method not correct

Cuff wrapping position on upper arm not at the same height as the heart

Patient's blood pressure not stable due to pulsus alternans, respiratory changes, or other reason

Incorrect cuff size used.

Measure circumference or patient's arm and ensure

correct sized cuff is used.

 

 

 

Cuff wrapped over thick clothing.

Ensure cuff is applied to a bare arm, or very thin

clothing.

 

 

 

Patient not seated properly.

Ensure patient is seated, feet flat on the floor, cuff at

heart level.

 

 

 

 

Ensure before measurement taken, patient has not

Patient ate, drank, or exerted themselves recently.

had food, caffeinated or alcohol beverages, or

 

exerted/exercised in the last 30 minutes.

Maintenance

27

Page 29
Image 29