Immunity test

IEC60601-1-2

IEC60601-1-2

 

 

Electromagnetic

test level

test level

 

environment - guidance

 

 

 

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any part

 

 

 

of the HBP-1300, including cables, than the

 

 

 

recommended separation distance calculated

 

 

 

from the equation applicable to the frequency of

 

 

 

the transmitter.

 

 

 

Recommend separation distance

 

 

 

d = 1.2

 

 

3 Vrms

 

P

 

 

 

d = 1.2

P

80 MHz to 800 MHz

Conducted RF

150 kHz to 80

3 Vrms

d = 2.3

P

800 MHz to 2.5 GHz

IEC 61000-4-6

MHz

 

 

 

 

 

80% AM (2Hz)

 

where P is the maximum output power rating

 

 

 

of the transmitter in watts (W) according to he

 

 

 

transmitter manufacturer and d is the recom-

 

3 Vrms

 

mended separation distance in meters (m).

 

 

 

 

 

Radiated RF

80 MHz to 2.5

3 V/m

Field strengths from fixed RF transmitters as

IEC 61000-4-3

GHz

 

determined by an electromagnetic site

 

80% AM (2Hz)

 

surveya, should be less than the compliance

 

 

 

level in each frequency rangeb.

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

 

 

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HBP-1300 is used exceeds the applicable RF compliance level above, the HBP-1300 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the HBP-1300.

bOver the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

34

Page 36
Image 36
Omron HBP-1300 specifications IEC60601-1-2 Electromagnetic Test level, Recommend separation distance