List of Error Codes

The alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority alarm occurs.

To clear an alarm, press any button.

If a low-priority alarm and a medium-priority alarm occur at the same time, the medium-priority alarm is displayed.

If the alarms are the same priority level, the error code of the alarm that occurred first is displayed. However, to prevent battery consumption when powered by battery, a low battery error (E40 error) is always given priority.

Example: E2

SYSTEM

Error

Priority

Description

Points to check

code

 

 

 

 

 

Sensor failure:

 

 

 

Sensor range not stable

 

 

 

Outside of sensor range

 

 

 

Pressure gain lower limit value / upper

 

 

 

limit value

 

E9

Medium

Pressure sensor count upper limit / lower

Contact OMRON HEALTHCARE Customer

limit error

Service.

 

 

 

 

 

EEPROM failure:

 

 

 

Checksum error

 

 

 

Read error

 

 

 

Protect code error

 

 

 

 

 

 

 

Switch error: Error in switch on state

 

 

 

 

 

NIBP

Error

Priority

Description

Points to check

code

 

 

 

 

 

The cuff tube is not connected

Firmly connect the cuff tube.

E1

Low

 

 

Air is leaking from the cuff.

Replace with an OMRON GS CUFF that

 

 

does not leak.

 

 

 

 

 

 

 

 

 

Did not inflate properly because the arm

Have the patient not move the arm or body,

 

 

or body moved during measurement.

and repeat measurement.

 

 

 

 

 

 

Moved body or arm during

Have the patient not talk or move, and

 

 

measurement, or talked.

repeat measurement.

 

 

 

 

 

 

The cuff is not applied correctly.

Correctly apply the cuff.

 

 

 

 

E2

Low

The sleeve is rolled up and is

Remove the garment and re-wrap the cuff.

compressing the arm.

 

 

 

 

 

 

 

 

 

 

 

The measurement time exceeds the

 

 

Measurement time has exceeded

expected time, so the measurement was

 

 

ended in order to avoid patient discomfort.

 

 

specified time.

 

 

There is a possibility that measurement is

 

 

Specified time: 165 seconds

 

 

being repeated over and over due to air

 

 

 

 

 

 

leaking from the cuff.

 

 

 

 