Power supply

AC adapter

Input voltage range: AC 100 V to 240 V

 

Rated Current: 1.6 A

 

Frequency: 50/60 Hz

 

Output voltage range: DC 6 V ±5%

 

 

Optional rechargeable battery

Type: 3.6 V, 1900 mAh

 

Number of operation cycles when fully charged: 300

 

Measurement conditions

 

• New battery fully charged

 

• Ambient temperature of 23°C (73.4°F)

 

• Using M-size cuff

 

• SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)

 

• One 5-minute cycle consisting of "cuff measurement time + wait time"

 

 

Environmental Conditions

Operational temperature and

Temperature range: 5 to 40°C (41 to 104°F)

humidity

Humidity range: 15 to 85%RH (not condensed)

 

Atmospheric pressure: 700 to 1060hPa

 

 

Storage and transportation

Temperature range: -20 to 60°C (-4 to 140°F)

 

Humidity range: 10 to 95%RH (not condensed)

 

Atmospheric pressure: 500 to 1060hPa

 

 

EMC: Reference standard

IEC60601-1-2:2007

 

Medical electrical equipment -

 

Part 1-2: General requirements for basic safety and essential performance

 

- Collateral standard: Electromagnetic compatibility - Requirements and

 

tests.

 

 

Non-Invasive Blood Pressure (NIBP)

Measurement technology

Oscillometric

 

 

Measurement method

Dynamic Linear Deflation method

 

 

Pressure display range

0 to 300 mmHg

 

 

Pressure display accuracy

Within ±3 mmHg or 2 %

 

 

NIBP measurement range

SYS 60 to 250 mmHg

 

DIA 40 to 200 mmHg

 

Pulse rate 40 to 200 /min

 

 

NIBP accuracy*

Maximum mean error within ±5mmHg

 

Maximum standard deviation within ±8mmHg

 

 

Pulse rate accuracy

Within ±5 % of reading

 

 

Reference standard:

IEC 80601-2-30:2009

 

ANSI/AAMI SP-10:2002+A1:2003+A2: 2006/(R)2008

 

ISO81060-2:2009

*Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is "K5".