Omron HBP-1300 specifications

Models: HBP-1300

1 40
Download 40 pages 60.07 Kb
Page 7
Image 7

Caution

Before using the unit, verify that none of the following apply to the patient:

-Poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be low blood flow to the measurement position)

-The patient uses an artificial heart and lung (there will be no pulse)

-An SpO2 sensor and the cuff are attached to the same arm

-The patient has an aneurysm

-The patient has frequent arrhythmia

-Body motions such as convulsions, arterial pulsations, or trembling (cardiac massage in progress, minute continuous vibrations, rheumatism, etc.)

Before use, visually inspect the unit to make sure there are no deformations due to falling, and that there is no dirt or moisture on the unit.

When the unit has not been used for an extended period of time, always verify that it operates normally and safely before use.

Do not use in a location where the unit may easily fall. In the event that the unit falls, verify that it operates normally and safely.

Introduction

5

Page 7
Image 7
Omron HBP-1300 specifications