Recommended Separation Distances:

Recommended separation distance between portable and mobile RF communications equipment and the HBP-1300

The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HBP-1300 as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

Rated maximum output

m

 

 

 

 

 

 

power of transmitter

 

 

 

 

 

 

 

 

 

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

(W)

 

 

 

 

 

 

 

 

 

 

d = 1.2

P

 

d = 1.2

P

 

d = 2.3

P

 

 

 

 

 

 

 

 

 

 

 

0.01

0.12

 

 

0.12

 

 

0.23

 

 

 

 

 

 

 

 

 

 

 

 

0.1

0.38

 

 

0.38

 

 

0.73

 

 

 

 

 

 

 

 

 

 

 

 

1

1.2

 

 

1.2

 

 

2.3

 

 

 

 

 

 

 

 

 

 

 

 

10

3.8

 

 

3.8

 

 

7.3

 

 

 

 

 

 

 

 

 

 

 

 

100

12

 

 

12

 

 

23

 

 

 

 

 

 

 

 

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note1:At 80MHz and 800MHz, the separation distance for the higher frequency range applies Note2:These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

36

Page 38
Image 38
Omron HBP-1300 specifications Recommended Separation Distances, Separation distance according to frequency of transmitter