7. Technical Data

7. Technical Data

Product Description

Digital Automatic Blood Pressure Monitor

Model

OMRON i-Q142 (HEM-1040-E)

Display

LCD Digital Display

Measurement Method

Oscillometric method

Measurement Range

Pressure: 0 mmHg to 299 mmHg

 

Pulse: 40 to 180/min.

Accuracy

Pressure: ￿3 mmHg

 

Pulse: ￿ 5% of display reading

Inflation

Fuzzy-logic controlled by electric pump

Deflation

Automatic pressure release valve

Pressure Detection

Capacitive pressure sensor

Memory

84 Measurements with date and time for each user

 

(A and B)

Power Source

AC/DC adapter (Included; Input 100-240V 50Hz/60Hz,

 

Output 6V/9.6W)

Operating temperature/

10°C to 40°C/30 to 85% RH

Humidity

 

Storage temperature/

-20°C to 60°C/10 to 95% RH

Humidity/Air Pressure

700 - 1060 hPa

Console Weight

Approximately 2600 g

Outer Dimensions

Approximately 286 (l) mm ￿ 294 (w) mm ￿ 271 (h) mm

Arm Circumference

22 to 42 cm

Package Content

Instruction manual, quick guide, AC adapter, guarantee

 

card, blood pressure pass and instruction manual for

 

optional printer

Note: Subject to technical modification without prior notice.

This OMRON product is produced under the strict quality system of OMRON Healthcare Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan for assembly.

= Type B

This device fulfils the previsions of EC directive 93/42/EEC (Medical Device Directive).

This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.