How to Select the Nebulization Mode

 

You can use this unit either in the continuous nebulization mode or in the manual mode.

Mode

Continuous Nebulization Mode

When you push the Button and immediately release

 

Nebulization

finger from the Button (within approximately

your finger

 

1 second), the unit nebulizes in Continuous

 

Nebulization Mode. When you push the Button

 

again, the unit stops nebulization.

 

Push the Button and

 

immediately release

Select the

(within 1 second).

Manual Nebulization Mode

to

When you keep pressing the Button (approximately

How

for 2 seconds or more), the unit nebulizes in Manual

 

 

Nebulization Mode. The unit nebulizes only while

Unit

you press the Button. You can inhale on demand.

Keep on pressing

the

the Button

of

(for approximately

2 seconds or more).

Use

– The power indicator (green light) illuminates during nebulization.

Correct

 

How to Inhale

1Hold the Main unit with your hand and start inhalation in

relaxed posture.

Slightly slant the Main Unit towards yourself.

Immerse the Vibrator in the medication. If the Vibrator is not immersed, the unit will not nebulize.

Medication

If saliva or medication collects on the electrode of Main Unit or on the Mesh, the unit stops nebulization. In that case, turn off the power and

absorb the medication or saliva with tissue paper or a soft cloth.

Caution

Do not drop or apply strong pressure to Main Unit, Medication Bottle Unit, or Mesh Cap.

• The unit may be damaged and cannot be used.

2When you finish inhalation, push the Button to turn off the

power.

•The unit has a built-in timer to turn off the power in approximately 30 minutes after the power is turned on.

•When an AC Adapter is used, unplug the power plug from the electric outlet. (Refer to Page 11)

Correct Use of the Unit How to Inhale

14

15

Page 8
Image 8
Omron NE-U22V instruction manual How to Select the Nebulization Mode, How to Inhale