Troubleshooting

If a problem occurs while you are using the unit, please check the following information before calling.

The nebulization rate is extremely low, or the unit does not nebulize at all.

Where to Inspect

How to correct

 

 

Replace the batteries with new ones.

Is the Battery Low Indicator turned on?

(Refer to Page 10)

Clean the Mesh Cap by disinfecting to remove

any stains.

(Refer to Page 18)

• If the trouble is not corrected by the above

Is the Mesh dirty?

mentioned procedure, replace the Mesh cap with a new one.

(Refer to Page 19)

The Power Indicator does not turn on and the unit does not nebulize.

Troubleshooting

 

Where to Inspect

How to correct

 

 

 

 

 

 

Are the polarities of batteries or rechargeable

Insert batteries or rechargeable batteries in the

 

 

 

 

correct alignment.

 

 

 

batteries placed in the wrong direction?

 

 

 

 

(Refer to Page 10)

 

 

 

 

 

 

 

 

 

 

Are the batteries worn?

Replace the batteries with new ones.

 

 

 

(Refer to Page 10)

 

 

 

 

 

 

 

 

 

 

Is the capacity of the rechargeable batteries

Recharge the batteries with the commercially

Warranty

 

 

available recharger or replace them

 

(NiMH) low?

 

 

 

 

with new ones.

 

 

 

 

 

 

 

 

 

 

 

 

Is the AC Adapter securely plugged into

 

 

and

 

the electric outlet or the connection port

Insert it correctly.

(Refer to Page 11)

 

of Main Unit?

 

 

 

 

 

 

Troubleshooting

 

 

 

 

 

Is the Medication Bottle installed to the Main

Install it correctly.

(Refer to Page 13)

 

 

 

 

Unit correctly?

 

 

 

 

 

 

 

 

 

Troubleshooting

The Power Indicator is illuminated, but the unit does not nebulize.

Where to Inspect

How to correct

 

 

 

 

 

 

 

 

Is the Medication Bottle filled with medication?

Fill with medication.

(Refer to Page 12)

 

 

 

 

 

 

Clean the Mesh Cap by disinfecting.

 

 

 

(Refer to Page 18)

 

Is the Mesh broken, stained,

• If the unit does not nebulize even after

 

carrying out the above-mentioned procedure,

 

or extremely soiled?

 

or the Mesh is broken, replace the Mesh Cap

 

 

 

 

with a new one.

 

 

 

 

 

(Refer to Page 19)

 

 

 

 

 

Is the medication being supplied to the

Use the unit slightly slanting the Main Unit

Troubleshootingand

Is the electrode of the Main Unit stained?

Wipe off the stain.

(Refer to Page 17)

Medication Bottle?

(the side with the Button) towards yourself.

 

(Is the medication in contact with the Vibrator?)

 

(Refer to Page 15)

 

Has any medication or water collected on the

Absorb the medication or water with

 

electrode of the Main Unit?

a soft cloth.

 

 

 

 

 

(Refer to Page 15)

 

 

 

 

 