Omron Healthcare HEM-907 How to Measure Blood Pressure in Single Mode, Set the Mode Selector to

Models: HEM-907

1 26
Download 26 pages 44.87 Kb
Page 16
Image 16

How to Measure Blood Pressure (in SINGLE Mode)

1. Wrap the arm cuff over the patient’s arm.

2. Push the ON/OFF (power)

Button to turn on the power.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

3. Set the MODE Selector to

“SINGLE”.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

4.Set the P-SET (pressure set- ting) Volume to “AUTO” or the target in ation value.

When setting the P-SET to “AUTO”, turn the Volume counterclockwise as far as it goes until you can hear the click sound.

When the systolic pressure is expected to exceed 220 mmHg, proper in ation may not be pos- sible in “AUTO” setting.

Set the target in ation value to 30 to 40 mmHg higher than the expected systolic pressure.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s

t/2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

5.Push the START Button to start the measurement.

If the in ation is determined insuf cient, the unit may start in ation again automatically.

If you want to stop measure- ment, push the STOP Button. The unit de ates fast.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

6.Measured results are dis- played.

SYS

mmHg

DIA

mmHg

PULSE

BPM

HIDE

7. Push the ON/OFF (power)

Button to turn off the power.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

16

Page 16
Image 16
Omron Healthcare HEM-907 instruction manual How to Measure Blood Pressure in Single Mode, Set the Mode Selector to