Omron Healthcare HEM-907 Troubleshooting, Problem What to inspect How to correct

Models: HEM-907

1 26
Download 26 pages 44.87 Kb
Page 23
Image 23

 

 

Troubleshooting

If a trouble occurs while using the unit, please check the following.

 

 

 

Problem

What to inspect

How to correct

 

Is the arm cuff wrapped correctly?

Wrap the arm cuff correctly, and measure

 

again.

 

 

 

 

 

 

Is the patient moving his/her arm or body

Ask the patient not to move his/her arm or

 

body during measurement, and measure

The unit in ates to abnormally

during in ation?

again.

high (low) pressure.

 

 

 

 

Does this trouble occur with a speci c

 

 

patient?

Check the measurement by the stetho-

 

For patients having arrhythmia, the unit

scope.

 

may not in ate properly

 

 

 

 

 

 

After checking the patient with the stetho-

 

First of all, check the patient’s condition.

scope as needed, take a proper mea-

 

sure in accordance with the “list of error

 

 

 

 

codes”.

 

 

 

 

Is the patient moving his/her arm or body

Ask the patient not to move his/her arm or

 

body during measurement, and measure

The unit cannot measure

during measurement?

again.

 

blood pressure.

 

 

Does the patient have an arrhythmia?

Check the patient with a stethoscope.

The blood pressure values are

extremely high (low).

 

 

Is the size of the arm cuff and the wrapping

Select the arm cuff in accordance to the

 

 

of arm cuff correct?

circumference of the patient’s arm, wrap it

 

correctly, then measure again.

 

 

 

 

 

 

Is the level of the brachium to which the

Keep the level of the brachium to which

 

arm cuff is wrapped at the same level as

the arm cuff is wrapped at the same level

 

the heart?

as the heart, then measure again.

 

 

 

23

Page 23
Image 23
Omron Healthcare HEM-907 instruction manual Troubleshooting, Problem What to inspect How to correct