Omron Healthcare HEM-907 instruction manual List of Error Codes, Error code Cause How to correct

Models: HEM-907

1 26
Download 26 pages 44.87 Kb
Page 22
Image 22

List of Error Codes

Error code

Cause

How to correct

Er1

In ation error

If the connector of rubber tube connecting the

 

When the pressure does not exceed 15

arm cuff and the main unit is loose, connect them

 

mmHg after operating a pump for 15 seconds

correctly, then measure.

 

from the start of in ation

If the rubber tube connecting the arm cuff and the

 

When the in ation does not reach the set

 

main unit is bent, stretch it, then measure.

 

pressure of the arm cuff within the speci ed

 

If the arm cuff is wrapped loosely, wrap it correctly,

 

time after starting the in ation

 

 

then measure.

Er2

De ation error

 

 

When the de ation speed is too fast during

 

 

the measurement

 

 

When the de ation speed is too slow during

 

 

the measurement

 

 

When the measurement does not end

 

 

within the speci ed time after starting the

 

 

measurement

 

 

When the pressure is applied more than 5

 

 

minutes

 

Er3

Overpressure error

If the rubber tube connecting the arm cuff and the

 

The arm cuff pressure exceeded 299 mmHg.

main unit is bent, stretch it, then measure.

Er4

Insuf cient in ation error

If the measurement is made by setting P-SET to

 

Blood pressure could not be measured due

“AUTO”, ask the patient not move his/her arm or

 

to insuf cient in ation after the measurement

body during the in ation.

 

ends.

When setting the P-SET to “AUTO”, turn the Volume

 

 

counterclockwise as far as it goes until you can hear

 

 

the click sound.

 

 

Set the P-SET to 30 to 40 mmHg higher than the

 

 

systolic pressure of the patient, then measure.

Er5

Indeterminable blood pressure error

If the arm cuff is wrapped loosely, wrap it correctly,

 

Blood pressure could not be measured even

then measure.

 

when the arm cuff pressure reached the

 

 

speci ed pressure.

 

Er6

Pulse wave small error

If the arm cuff is wrapped loosely, wrap it correctly,

 

Pulse wave was too small.

then measure.

 

 

Do not wrap the arm cuff over a thick cloth or by

 

 

rolling up the jacket or shirt.

Er7

Measurement error

Check if the patient has an arrhythmia.

 

Relationship between systolic and diastolic

Ask the patient not to move his/her arm or body

 

pressures was abnormal.

 

during the measurement.

Er8

Pulse rate error

 

 

Pule rate did not stay within the range of 30

 

 

to 199 beats/min.

 

Er9

Device error

Contact the dealer where you purchased the unit or

 

There is a problem with the main unit of the

the nearest OMRON dealer.

 

blood pressure monitor.

 

22

Page 22
Image 22
Omron Healthcare HEM-907 instruction manual List of Error Codes, Error code Cause How to correct