Omron Healthcare HEM-907 instruction manual How to Measure Blood Pressure in MANU. Mode

Models: HEM-907

1 26
Download 26 pages 44.87 Kb
Page 18
Image 18

How to Measure Blood Pressure (in MANU. Mode)

1.Wrap the arm cuff over the patient’s arm.

3.Push the ON/OFF (power) Button to turn on the power.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

2.Place the stethoscope on the patient’s arm.

4.Set the MODE Selector to “MANU.”.

ON/OFF

START

HIDE

 

 

DEFL ATION

P-SET

AVG./1s

t/2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

5.Set the P-SET (pressure set- ting) Volume to “AUTO” or the target in ation value.

When setting the P-SET to “AUTO”, turn the Volume coun- terclockwise as far as it goes un- til you can hear the click sound.

When the systolic pressure is expected to exceed 220 mmHg, proper in ation may not be pos- sible in “AUTO” setting.

Set the target in ation value to 30 to 40 mmHg higher than the expected systolic pressure.

ON/OFF

START

HIDE

 

 

DEFL ATION

P-SET

AVG./1s t/2nd/3rd

MODE

140

180

SINGLE

 

AVG.

MANU.

100

220

CHECK

STOP

 

260

AUTO

280

6.Push the START Button to start the measurement.

In the “AUTO” setting, the unit stops in ation at the pres- sure considered to be 30 to

40 mmHg above the expected systolic pressure, then start de ation.

The unit can be re-in ated

only while the START Button is pushed.

The pressure display is syn- chronized with the pulse rate and renewed.

Each time the DEFLATION (de ation control) Button is pushed, the unit de ates by 5~10 mmHg.

When you press the STOP But- ton, the unit de ates fast and ends the operation.

ON/OFF

START

HIDE

 

 

 

DEFL ATION

P-SET

 

AVG./1s t/

2nd/3rd

MODE

 

140

180

SINGLE

 

 

AVG.

MANU.

 

100

220

CHECK

 

STOP

 

260

AUTO

280

7. Push the ON/OFF (power)

Button to turn off the power.

 

ON/OFF

 

 

 

HIDE

START

 

 

 

 

DEFL ATION

P-SET

MODE

AVG./1s t/2nd/3rd

 

140 180

SINGLE

MANU.

 

AVG.

100

220

CHECK

 

260

STOP

 

 

AUTO

280

 

18

Page 18
Image 18
Omron Healthcare HEM-907 instruction manual How to Measure Blood Pressure in MANU. Mode