How to Clean the Unit after Use

Caution

Pull the AC adapter from the electric outlet when cleaning the unit.

You may suffer electric shock.

After cleaning the unit, dry it well, then insert the AC adapter to the electric outlet.

You may suffer electric shock.

1.Clean the blood pressure monitor,the cuff and the cables using a cloth dampened with water,diluted disinfectant alcohol or diluted mild detergent.

To disinfect, wipe the unit with a soft cloth squeezed well after moistened with the following disinfectant.

Benzalkonium chloride

0.01 - 0.2 w/v%

Chlorhexizin

0.05 - 0.5 w/v%

Amphoteric surface active agent

0.01 - 0.2 w/v%

When cleaning or disinfecting the unit, do not wipe the AC adapter.

2.Then wipe the unit with a soft dry cloth.

Requests from OMRON

Do not clean the unit with gasoline, thinner, or high concentration alcohol.

Do not sterilize the unit by autoclave or gas sterilization (EOG, formaldehyde, or high concentration ozone.)