How to Measure Blood Pressure (in CHECK Mode)
Display of pressure can be con rmed by the CHECK Mode (yearly procedure).
What you need to prepare
(1)
(2)
(3)Two rubber tubes, and (4) something in the cylindrical shape to wrap the arm cuff such as the tea container (sturdy one that will not break or deform with pressure)
*Measurement error may occur even with the mercury blood pressure monitor because of insuf cient amount of mercury in the mercury, dirty glass tube, or clogged glass tube cap.
How to use the blood pressure monitor
1.Connect the mercury blood pressure monitor, rubber ball, and this unit with the
2.Push the ON/OFF (power) Button to turn on the power.
3.Set the MODE Selector to “CHECK”.
4.Tightly wrap the arm cuff over a sturdy cylindrical object.
5.Close the air release valve of rubber ball and in ate the monitor to a certain value (pressure to be checked).
|
|
| SYS |
|
|
|
|
|
| mmHg | |
|
|
| DIA |
|
|
|
| P | ULSE | mmHg | |
|
|
|
|
| |
|
|
|
| B |
|
|
|
|
| PM AC | |
|
|
|
|
| HIDE |
| P- |
|
|
| |
| 140 | SET |
|
| |
|
| 180 |
| MODE | |
100 |
|
|
| ||
|
|
|
| ||
|
|
| 220 | AVG. | SINGLE |
|
|
|
|
| MANU. |
AUTO |
|
| 260 |
| CHECK |
|
|
| 280 |
|
|
| ON/OFF |
| ST |
| ART |
A | DEF |
VG./1 LATION | |
| st /2nd/3rd |
| STOP |
6.Compare the value displayed on this monitor and that on the mercury.
7.Open the air release valve of rubber ball to de ate.
CHECK result
(1)In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the factory) should not exceed 3 mmHg.
(2)If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON dealer.
19