DIAGNOSTIC PROCEDURES

USB Video Diagnostic - If live video is not displayed on the PC monitor of the Polaris System when the imaging software application is running, there could be a problem with the settings in the software application or with the handpiece. Perform the following USB Video Diagnostic procedure to check that the Polaris handpiece is working correctly, the Polaris Handpiece Cable is correctly connected between the Polaris System and the computer and the

Polaris System USB driver files are installed correctly on the PC.

1.Go to the Start button and select Programs.

2.Select the Polaris program option and click on Camera Demostration Program and observe that the Camera Demonstration Program screen opens.

3.Depending on the outcome of the diagnostic, perform one of the following:

a.If a live image is correctly displayed, exit from the application and make sure that the imaging software application is configured correctly.

b.If a live image is not displayed, exit from the application and check the handpiece, Polaris Handpiece Cable and USB driver file installation.

4.Reset the driver's video settings by performing the following:

a.On the Polaris Demostration Program screen, click the Properties button located on the lower left of the live image screen.

b.On the Properties screen select the VIDEO PROC AMP tab and then click the DEFAULT button. The image colors should now be correct. Click OK to close the Properties screen and exit from application.

Live

Intraoral Camera Demonstration Program - Polaris Intraoral Video Camera Image

Screen

Properties

Properties

Screen

Video

Proc Amp

Screen

Properties

Tab Opens Video

Proc Amp Screen

Properties

 

Default Button: Selects

Button:Opens

 

 

Default Settings as Shown

Properties Screen

 

 

 

OK Button: Closes

 

Properties Screen

 