Operator Manual, Supercharger 60 R3.0
Page 46 of 62
Set the Cell Selector at 20, connect a voltage source set to
approximately 21V and start the Charger-Analyzer in AUTO
Discharge
Adjust for a Discharge CUT-OFF (Capacity Failure) at 20.0 volts.
NOTE: This step requires that the Overvoltage cut-off on the
MONITOR circuit board be previously calibrated.
9.2.3.3. CLOCK-TIMER
No adjustments on this board.
Older versions of the Charger-Analyzer use two circuit boards
associated with timer functions: The Timer Board 9879903010 and
the Clock Display Board 9879903009.
Newer Superserseders use a single Clock-Timer Board
9879903109. In either case, there are no adjustments to perform.
9.2.3.4. (Blank)
9.2.3.5. CURRENT CONTROL (see fig 11).
NOTE: Use the hook at pin 14 as ground reference for voltage
measurements (test point E0).
9.2.3.5.1. R29 - 10 VOLT ADJUSTMENT.
Static test (no battery or Calibrator).
Adjust for a reading of 10.000V ±0.003V at E7.
NOTE: Newer versions of this circuit board have a
fixed 10V reference (no adjustment). Verify that it
reads 10V ±0.005V.
9.2.3.5.2. R28 - LOW CURRENT CALIBRATE.
Connect a battery and start the Charger-Analyzer in
Single Rate.
Adjust for a reading of 0.9 to 1.0 AMP at the current
digital panel meter, with the topping charge current
potentiometer set for 1 AMP (010).
9.2.3.5.3. R25 - HIGH CURRENT CALIBRATE
Connect a battery and start the Charger-Analyzer in
Dual Rate.
Adjust for a reading of 25.0 AMPS at the current
digital panel meter, with the main current
potentiometer set for 25.0 AMPS (250).