DAILY CLEANING OF THE MASK (OR AS REQUIRED)

Disconnect the main air tubing from the swivel.

The RESMED HOSPITAL NASAL MASK is for single patient, multi-use. The mask may be cleaned using a 70% v/v isopropyl alcohol swab or by wiping with a damp, soapy cloth. If using a damp cloth, use mild soap or diluted dishwashing detergent. Allow to air dry out of direct sunlight.

Note: Do not disassemble the mask.

CAUTIONS

Do not use solutions containing bleach, chlorine, aromatics, moisturizers, antibacterial agents, or scented oils to clean any part of the system or air tubing. These solutions may cause damage and reduce the life of the product.

Do not expose any part of the system or tubing to direct sunlight as it may deteriorate.

If any visible deterioration of a component is apparent (cracking, crazing, tears, etc), replace the mask.

Check the vent holes for blockage. Ensure the vent holes are cleaned regularly.

TECHNICAL SPECIFICATIONS

PRESSURE / FLOW CURVE

Note: As a result of manufacturing variations, the vent flow may vary from the values listed in the following table.

Pressure

Flow

 

50.0

 

 

 

 

 

 

 

 

(cm H2O)

(L/min)

 

45.0

 

 

 

 

 

 

 

 

4

19

 

40.0

 

 

 

 

 

 

 

 

(L/min)

35.0

 

 

 

 

 

 

 

 

8

28

30.0

 

 

 

 

 

 

 

 

Rate

25.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

12

34

Flow

20.0

 

 

 

 

 

 

 

 

 

15.0

 

 

 

 

 

 

 

 

16

40

 

10.0

 

 

 

 

 

 

 

 

20

45

 

5.0

 

 

 

 

 

 

 

 

 

0.0

 

 

 

 

 

 

 

 

 

 

 

4

6

8

10

12

14

16

18

20

 

 

 

 

 

 

 

Pressure (cm H20)

 

 

 

 

FLOW GENERATOR SETTING

Select “Standard” as the mask option, when using the RESMED HOSPITAL NASAL MASK with ResMed flow generators that have mask setting options.

DEAD SPACE INFORMATION

Explanation: Dead space is the empty volume of the mask up to the swivel.

The dead space of the mask varies according to cushion sizes, but is less than 104 mL.

ENVIRONMENTAL CONDITIONS

Operating Temperature: 41°F (5°C) to 104°F (40°C)

Operating Humidity: 15–95% relative humidity

ENGLISH

Page 3
Image 3