Problem | Possible Cause | Solution |
|
|
|
| Breath is not deep | Take a deep breath in and |
| enough to trigger | out through the mask. |
| SmartStart. |
|
| There is excessive leak. | Adjust position of mask |
|
| and headgear. |
|
| Plugs may be missing |
|
| from ports on mask. |
|
| Replace them. |
|
| Air tubing not connected |
|
| properly. Connect firmly at |
|
| both ends. |
|
| Air tubing kinked or |
|
| punctured. Straighten or |
|
| replace. |
|
|
|
VPAP unit does not | SmartStart/Stop is | Enable SmartStart/Stop. |
stop when you | disabled. |
|
remove your mask. |
|
|
| Use of an Ultra Mirage | SmartStop does not work |
| Full Face Mask. | with a full face mask. |
|
|
|
SmartStart is enabled | Incompatible humidifier or | Use only equipment as |
but the flow | mask system being used. | recommended and |
generator does not |
| supplied by ResMed. |
stop automatically |
|
|
when you remove |
|
|
your mask. |
|
|
Display error | The air tubing is loose. |
message: |
|
Check tube!! |
|
Key if done |
|
Check that the air tubing is connected securely to your mask and the air outlet on the front of the VPAP.
To clear the error message, press any key on the VPAP keypad.
Displays error | Component failure. | Return your VPAP for |
message: |
| servicing. |
SYSTEM ERROR |
|
|
Call service! |
|
|
|
|
|