Function

Default

Function Description

Setting Options

 

 

 

 

Humidifier

NONE

Selects the type of humidifier to

NONE, PASSOVER,

 

 

be used with the VPAP.

HUMIDAIRE

 

 

 

If the HumidAire 2i

 

 

 

is used, the above

 

 

 

setting options do

 

 

 

not appear and H2i

 

 

 

is displayed.

 

 

 

 

SmartStart

OFF

Turns the SmartStart function on

ON/OFF

 

 

or off.

 

 

 

See “SmartStart™” on page 23.

 

 

 

 

 

Mask Alarm

ON

Turns the Mask Alarm signal on

ON/OFF

 

(VPAP III ST);

or off.

 

 

OFF

If enabled, the Mask Alarm

 

 

(VPAP III)

feature will alert you when a

 

 

 

high mask leak is detected. An

 

 

 

audible tone will sound and a

 

 

 

high leak message will appear

 

 

 

on the LCD screen.

 

SmartStart™

The VPAP has a function called SmartStart which can be enabled by your clinician. If SmartStart is enabled, VPAP will start automatically when you breathe into the mask and will stop automatically when you take your mask off. This means you do not have to press the Front key to begin or end treatment.

Note: If you select “Mir Full” as the mask option, SmartStop is automatically disabled. SmartStart may not work with an Ultra Mirage Full Face Mask due to safety features of the mask.

When Mask Alarm is set to ON, SmartStart/Stop automatically reverts to OFF. SmartStop cannot be used with Mask Alarm because if a high leak occurs, SmartStop will stop treatment before the Mask Alarm signal is activated.

Settings for Mask Types

The following table shows the setting that should be selected for each mask type.

Table 3: Settings for mask types

Settings

Mask

 

 

MIRAGE

Mirage Nasal Mask

 

 

ULTRA

Ultra Mirage Nasal Mask

 

 

How to Use the VPAP Menus

23

Page 30
Image 30