System Specifications
Dynamic pressure characteristics
IPAP: 2 cm H2O to 25 cm H2O (measured at the end of standard 2 m air tubing)
EPAP: 2 cm H2O to 25 cm H2O (measured at the end of standard 2 m air tubing)
CPAP: 4 to 20 cm H2O (measured at the end of standard 2 m air tubing)
Maximum single fault pressure: 40 cm H2O
Maximum flow (CPAP mode, measured at the end of standard 2 m air tubing)
Pressure | Flow |
(cm H2O) | (L/min) |
10 | 200 |
15170
20150
Sound pressure level: <30 dB (tested in accordance with the requirements of ISO
Dimensions (L x W x H): 270 mm x 230 mm x 141 mm
Weight: 2.3 kg
Air outlet: 22 mm taper, compatible with EN
Pressure measurement: Internally mounted pressure transducer
Flow measurement: Internally mounted flow transducer
Power Supply: Input range
Housing Construction: Flame retardant engineering thermoplastic
Environmental Conditions
Operating Temperature: +5°C to +40°C
Operating Humidity:
Storage and Transport Temperature:
Storage and Transport Humidity:
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements (EMC) according to
Air Filter:
System Specifications | 35 |