ResMed s8 manual Technical Specifications, Symbols which appear on the device, English

Models: s8

1 88
Download 88 pages 45.08 Kb
Page 19
Image 19
Technical Specifications

Technical Specifications

Operating pressure range

4 to 20 cm H2O

Dimensions (H x W x D)

4.6" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)

Weight

3.1 lb (1.4 kg)

Operating Temperature

+41°F to +95°F (+5°C to +36°C)

Power Supply

• Input range for S8 Series II with HumidAire 3i: 100–240V, 50/60Hz;

 

110V, 400Hz; 2.5A <140VA (110W) (maximum power consumption).

 

Instantaneous peak power consumption <340VA.

 

• Refer to the DC-12 converter instructions for DC ratings. Only use

 

the ResMed DC-12 converter for DC input.

 

• Typical power consumption at 20 cm H2O is 39VA or 20.7W.

Operating Humidity

10%–95% non-condensing

Storage and Transport

-5°F to -140°F (-20°C to +60°C)

Temperature

 

Storage and Transport

10%–95% non-condensing

Humidity

 

Operating Altitude

Sea level to 8500’ (2591 m)

Electromagnetic

Product complies with all applicable electromagnetic compatibility

Compatibility

requirements (EMC) according to IEC60601-1-2, for residential,

 

commercial and light industry environments. The electromagnetic

 

compatibility tables for these ResMed devices can be found on

 

www.resmed.com, on the Products page under Service and

 

Support. Click on the PDF file for your device.

Air Filter

Two-layered, powder-bonded, polyester non-woven fibre

Air Tubing

Flexible plastic, 6’ 6" (2 m)

Air Outlet

The 22 mm conical air outlet complies with EN 1281-1

IEC 60601-1 Classifications

Class II (double insulation), Type CF

Supplemental Oxygen

Recommended maximum supplemental oxygen flow: 4 L/min

Housing Construction

Flame retardant engineering thermoplastic

Noise Level

Radiated sound pressure is measured at <24±2 dBA at 10 cm H2O and

 

3’3.36” (1 m) according to ISO17510-1

The manufacturer reserves the right to change these specifications without notice.

Symbols which appear on the device

Attention, consult accompanying documents;

Drip proof;

Type CF equipment;

Operating Altitude Dangerous voltage; Supplemental OxygenHousing ConstructionSymbols which appear on the device Class II equipment; Manual background Start/Stop.

English

Troubleshooting 15

Page 19
Image 19
ResMed s8 Technical Specifications, Symbols which appear on the device, English, Operating Altitude, Supplemental Oxygen