14

Tailoring treatment setup options

Setting Ramp

Ramp time is a feature that can be enabled
by your clinician by setting a maximum ramp
time. Designed to make the beginning of
treatment more comfortable, ramp time is
the period during which the pressure
increases from a low start pressure to the
treatment pressure. See “Setup menu:
Options” on page 16.

Programs

If the Clinician has selected both, ie, dual,
you can choose the program to use on the
Treatment screen. If only a single program is
selected, the option does not display.
Using mask-fit
You can use mask-fit to help you fit your mask properly. This feature delivers constant
treatment pressure for a three-minute period, prior to starting treatment, during which you
can check and adjust your mask-fit to minimise leaks. The mask-fit pressure is the set
CPAP or EPAP pressure or 10 cm H2O, whichever is greater.
1Put the mask on according to the mask user guide.
2Hold down for at least three seconds until pressure delivery starts.

3

If necessary adjust the mask, mask cushion and headgear until you have a good mask fit.
After three minutes treatment will begin. Mask-fit can be stopped at any time by pressing .
Notes:
Press for at least three seconds during mask-fit to start the treatment immediately.
The mask-fitting feature is disabled when the mask type Trac h is selected.
Ramp time
Program