Problem / possible cause

Solution

FiO2 sensor is used or defective.

If the lifetime of the FiO2 sensor has

 

exceeded one year please replace the

 

FiO2 sensor and start calibration again.

LCD: is displayed in the header.

 

 

Battery is not charging.

• Ensure the ambient temperature

 

 

 

conditions are within the specified

 

 

 

operating range. If the problem persists

 

 

 

within the specified operating conditions

 

 

 

please return the device for servicing.

 

 

 

• Power off the device. Power son the

 

 

 

device again.

 

 

 

Technical specifications

 

 

 

 

 

 

 

Operating pressure

• IPAP: 2 cm H2O to 40 cm H2O (in S, ST, T, PAC mode)

 

 

range

• PS: 0 cm H2O to 38 cm H2O (in S, ST, T, PAC mode)

 

 

 

• EPAP: 2 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC

 

 

 

mode)

 

 

 

• CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only)

 

 

 

• Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode)

 

 

 

• Max PS: 0 cm H2O to 30 cm H2O (in iVAPS mode)

 

 

 

Note: iVAPS mode is only available in Stellar 150.

 

 

 

 

 

 

Maximum single fault

60 cm H2O (in all modes)

 

 

 

pressure

 

 

 

 

 

 

 

 

 

Maximum breathing

2 cm H2O at 30 L/min;

 

 

resistance under single

7.2 cm H2O at 60 L/min

 

 

fault

 

 

 

 

 

 

 

 

 

 

Maximum flow

> 200 L/min at 20 cm H2O

 

 

 

Flow accuracy

± 5 L/min or 20% measured value, whichever is greater

 

 

 

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

 

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

 

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

 

 

 

 

Therapy pressure

IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of

 

 

 

tolerance

inspiration)

 

 

 

EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure

 

 

 

CPAP: ± 0.5 cm H2O ± 10% of set pressure

 

 

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

 

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

 

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

 

 

 

 

Sound pressure level

29 dBA as measured according to ISO 17510 – 1; 2002.

 

 

 

 

32 dBA with uncertainty of 3 dBA as measured according

 

 

 

to ISO 17510 – 1:2007.

 

 

 

 

 

 

 

Alarm Volume Range

> 45 dBA - <85 dBA at 1 meter (3 steps: low, medium, high)

 

 

 

 

 

 

 

Dimensions (L x W x H)

230 mm x 170 mm x 120 mm

 

 

 

 

 

 

 

Weight

2.1 kg

 

 

 

 

 

 

 

Air outlet

22 mm taper, compatible with ISO 5356-1:2004

 

 

 

Anaesthetic & Respiratory Equipment - Conical

 

 

 

Connectors

 

 

 

 

 

 

 

Pressure measurement

Internally mounted pressure transducer

 

 

 

 

 

 

 

Flow measurement

Internally mounted flow transducer

 

 

 

 

 

English

Technical specifications

27

Page 31
Image 31
ResMed Stellar 100 and 150, Stellar 150 manual Technical specifications