Problem / possible cause | Solution |
FiO2 sensor is used or defective. | If the lifetime of the FiO2 sensor has |
| exceeded one year please replace the |
| FiO2 sensor and start calibration again. |
LCD: is displayed in the header.
|
| Battery is not charging. | • Ensure the ambient temperature | |
|
|
| conditions are within the specified | |
|
|
| operating range. If the problem persists | |
|
|
| within the specified operating conditions | |
|
|
| please return the device for servicing. | |
|
|
| • Power off the device. Power son the | |
|
|
| device again. | |
|
|
| ||
Technical specifications | ||||
|
|
|
|
|
|
| Operating pressure | • IPAP: 2 cm H2O to 40 cm H2O (in S, ST, T, PAC mode) | |
|
| range | • PS: 0 cm H2O to 38 cm H2O (in S, ST, T, PAC mode) | |
|
|
| • EPAP: 2 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC | |
|
|
| mode) | |
|
|
| • CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only) | |
|
|
| • Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode) | |
|
|
| • Max PS: 0 cm H2O to 30 cm H2O (in iVAPS mode) | |
|
|
| Note: iVAPS mode is only available in Stellar 150. | |
|
|
|
| |
|
| Maximum single fault | 60 cm H2O (in all modes) |
|
|
| pressure |
|
|
|
|
|
|
|
|
| Maximum breathing | 2 cm H2O at 30 L/min; | |
|
| resistance under single | 7.2 cm H2O at 60 L/min | |
|
| fault | ||
|
|
|
| |
|
|
|
| |
|
| Maximum flow | > 200 L/min at 20 cm H2O |
|
|
| Flow accuracy | ± 5 L/min or 20% measured value, whichever is greater |
|
|
|
| Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm | |
|
|
| H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec, | |
|
|
| Respiratory Rate: 10 bpm, with ResMed calibration cap. | |
|
|
|
| |
|
| Therapy pressure | IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of |
|
|
| tolerance | inspiration) | |
|
|
| EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure | |
|
|
| CPAP: ± 0.5 cm H2O ± 10% of set pressure | |
|
|
| Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm | |
|
|
| H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec, | |
|
|
| Respiratory Rate: 10 bpm, with ResMed calibration cap. | |
|
|
|
| |
|
| Sound pressure level | 29 dBA as measured according to ISO 17510 – 1; 2002. |
|
|
|
| 32 dBA with uncertainty of 3 dBA as measured according | |
|
|
| to ISO 17510 – 1:2007. | |
|
|
|
|
|
|
| Alarm Volume Range | > 45 dBA - <85 dBA at 1 meter (3 steps: low, medium, high) | |
|
|
|
|
|
|
| Dimensions (L x W x H) | 230 mm x 170 mm x 120 mm | |
|
|
|
|
|
|
| Weight | 2.1 kg | |
|
|
|
|
|
|
| Air outlet | 22 mm taper, compatible with ISO | |
|
|
| Anaesthetic & Respiratory Equipment - Conical | |
|
|
| Connectors | |
|
|
|
|
|
|
| Pressure measurement | Internally mounted pressure transducer | |
|
|
|
|
|
|
| Flow measurement | Internally mounted flow transducer | |
|
|
|
|
|
English |
Technical specifications | 27 |