Problem / possible cause

Solution

 

 

There is a large impedance (eg, antibacterial filter) in the air circuit.

Perform the Learn Circuit function.

The device does not stop when you remove your mask

SmartStart/Stop is disabled.

Consult your clinician.

Use of a full face mask or tracheostomy

SmartStart is disabled if Full face mask or

tube.

Trach is selected as interface.

 

 

Incompatible accessories (eg, humidifier

Use only equipment as recommended

or mask system) with high resistance

and supplied by ResMed.

being used.

 

 

 

High Leak Alarm or Low Min Vent alarms

Consult your clinician.

are set to ON.

 

 

 

Confirm Stopis enabled.

Consult your clinician.

High Leak Alarm is enabled, but alarm does not activate when the mask is removed during treatment

Incompatible air delivery system being

Use only equipment as recommended

used.

and supplied by ResMed.

 

 

Pressure settings are too low for the air

Perform the Learn Circuit function to

delivery components being used.

adjust the therapy pressure according to

 

your air tubing system.

 

 

Learn Circuit failed

 

The circuit configuration is inappropriate as the impedance detected is too high.

Too many components have been included or the impedance of accessories in use is above ResMed’s recommendation, eg, type of filter, external humidifier, air tubing.

Review the components included in the circuit configuration and adjust as appropriate, then rerun Learn Circuit (see “Setup menu” on page 16).

The delivered airflow is not humid/heated although the H4i humidifier is in use

The humidifier is not properly attached.

Correctly attach the humidifier.

The humidifier does not heat.

The device is currently powered by

 

battery use or not connected to the

 

mains.

The humidifier does not work.

Return the device and the humidifier for servicing.

The water chamber is empty.

Fill the water chamber of the humidifier.

USB stick is not readable or writeable

 

 

The USB stick contains unreadable data,

Consult your clinician.

does not have enough space available, is

 

not compatible with the device.

 

 

 

USB stick is defective.

Replace the USB stick after consulting

 

your clinician.

FiO2 sensor calibration failure

FiO2 sensor is not attached properly.

For the correct attachment of the FiO2 sensor, see “Using the FiO2 monitoring sensor” on page 8.

26

Page 30
Image 30
ResMed Stellar 100 and 150, Stellar 150 Problem / possible cause Solution, Device does not stop when you remove your mask