.

Air filter

Air filter cover

English

Disinfection

Disinfection of your device helps to prevent the risk of cross contamination.

Disinfect the exterior of the device, and especially the air outlet, with a damp cloth and a disinfection solution (eg, Microzid).

Multipatient use

WARNING

• An antibacterial filter is mandatory if the device is used on multiple patients.

In a mulitpatient use environment, you must perform the following before the device is provided to a new patient:

Air filter and the

Replace.

antibacterial filter

 

Mask

Reprocess; Cleaning, disinfection and sterilisation

 

instructions are available from the ResMed website,

 

www.resmed.com/masks/sterilization. If you do not have

 

Internet access, please contact your ResMed representative.

Air tubing

Replace the air tubing. Alternatively, consult the air tubing

 

instructions for cleaning and disinfection information.

Device

Disinfect the Stellar as follows:

 

Use an anti-bacterial cleaning/disinfectant solution such as

 

Mikrozid® AF or CaviCide® with a clean non-dyed

 

disposable cloth to clean and disinfect the exterior surfaces

 

of the device. Wipe all accessible surfaces of the device,

 

including the air outlet (avoid liquid entering any openings in

 

the device). Follow the manufacturer’s recommended

 

cleaning instructions.

Humidifier

As instructions for humidifiers vary, see the user guide for

 

the humidifier in use. For multipatient use environment, use

 

the H4i water chamber (disposable) in place of the H4i

 

reusable water chamber.

Cleaning and maintenance

21

Page 25
Image 25
ResMed Stellar 100 and 150, Stellar 150 manual Disinfection, Multipatient use