Power supply

AC 100–240V, 50–60Hz, 2.2 A, max. 65 W

External DC Power

24 V, 3 A

Supply (isolated)

 

 

 

 

Internal Battery

Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh

 

Operating hours: 2 h with a new battery under normal

 

conditions (see below).

 

Patient type: home chronic; pressure: IPAP/EPAP 15/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 20 bpm; battery capacity: 100%

 

Patient type: hospital acute; pressure: IPAP/EPAP 20/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 45 bpm; battery capacity: 100%

 

 

Housing construction

Flame retardant engineering thermoplastic

 

 

Environmental

• Operating temperature: 0°C to 35°C

conditions

• Operating humidity: 10%–95% non-condensing

 

• Storage and transport temperature: -20°C to 60°C

 

 

(+50°C*)

 

• Storage and transport humidity: 10%–95% non-

 

 

condensing

 

• Air pressure: 680hPa to 1,100hPa; Altitude: 3,000 m

 

*NONIN XPOD

 

 

Electromagnetic

Product complies with all applicable electromagnetic

compatibility

compatibility requirements (EMC) according to

 

IEC60601-1-2, for residential, commercial, and light

 

industry environments. For further details see “Guidance

 

and manufacturer’s declaration – electromagnetic

 

emissions and immunity” on page 29.

 

 

Air filter

Electro static fibre mesh with TPE frame structure. Bacterial

 

filtration efficiency of 99.540% on area weight 100g/m².

 

 

Air tubing

Flexible plastic, 2 m or 3 m length (22 mm diameter)

 

 

SlimLine air tubing

Flexible plastic, 1.83 m length (15 mm diameter)

 

 

IEC 60601-1

• Class II (Clause 3.14—double insulation). This

classifications

 

adherence means the need for an protective earthing

 

(ie, an earthed plug) is not necessary.

 

Type BF

 

Continuous operation

Air travel requirements Medical-Portable Electronic Devices (M-PED) that meet the Federal Aviation Administration (FAA) requirements of RTCA/DO-160 can be used during all phases of air travel without further testing or approval by the airline operator. ResMed confirms that the Stellar meets RTCA/ DO-160 requirements.

This device is not suitable for use in the presence of a flammable anesthetic mixture.

Notes:

The manufacturer reserves the right to change these specifications without notice.

Pressure may be displayed in cm H2O or hPa.

28