Troubleshooting and Displays

The thermometer has been designed to display feedback/warnings to ensure proper usage and accurate temperature measurement. When an error or indication appears, one “beep” signal will be heard once every 5 seconds for 10 seconds and then the thermometer will turn off automatically. In the event of a malfunction or incorrect temperature measurement, the following warnings will be displayed:

Display

Meaning

Action

 

The measured

Turn the unit OFF and then ON

 

temperature is below

again. Retake temperature as

 

95°F (35°C).

described in “Using Your

 

 

Thermometer” section.

 

The measured temperature

Turn the unit OFF and then ON

 

is above 107.6°F (42°C).

again. To verify the reading, retake

 

Or thermometer was

temperature as described in “Using

 

placed in body position

Your Thermometer” section.

 

before two beeps were

 

 

heard.

 

 

The thermometer is too

Verify room temperature is within

 

cold or too hot to

50°F to 104°F (16°C to 40°C).

 

operate.

Retake temperature as described in

 

 

“Using Your Thermometer” section.

 

Temperature did not

Turn the unit OFF, wait two

 

register because of poor

minutes, turn it ON again and

 

contact, patient movement

retake temperature as described in

 

or poor placement.

“Using Your Thermometer” section.

 

Unit malfunctioned.

Do not use the thermometer –

 

 

contact distributor for instructions.

 

Battery icon is on

The batteries have enough

 

continuously.

remaining life for a few more

 

 

measurements. Replace batteries.

 

Only battery icon is on

The batteries are drained - replace

 

and the device

batteries (see “Replacing Batteries”

 

automatically turns

section).

 

itself OFF.

 

 

 