SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

The SENSOR FAILURE message for toxic gas sensors does not indicate a specific failure of the sensor but indicates that the sensor is not correctly connected to the electronics. If this message appears check that the sensor harness is correctly installed to the connector on the bottom electronics board.

The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. To replace the sensor:

1.Confirm that system power has been removed.

2.Remove the transmitter electronics board from the main housing and unplug the sensor harness from the transmitter electronics.

3.Unscrew the sensor housing from the bottom of the enclosure

4.Hold the sensor assembly so that the harness faces down and the sensor faces up. Unscrew and remove the round section of the housing from the hex

 

section. Be careful not to lose the spacer

 

 

 

Figure 7-12

 

 

 

 

 

 

 

 

washer which will be sitting on top of the

 

Toxic Gas Transmitter Component Locator

 

 

 

 

 

 

exposed sensor.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5.

Carefully pull the old sensor straight up from the

1.

Adjust

the

span

potentiometer so

that

it

is

 

 

approximately

mid

range.

(Twenty

turns

 

 

socket.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

clockwise and tens turns counter-clockwise).

 

 

 

 

6.

Orient the

new sensor

so that

the

sensor pin

 

 

 

 

 

2.

Remove system

power and remove

both

the

 

 

labeled “C” faces the socket labeled “C” which is

 

 

 

transmitter

and

the sensor assembly

from

the

 

 

on the far side of the board from the vertical reed

 

 

 

 

enclosure. Remove the sensor assembly cover so

 

 

switch. The reed switch will slide into a hole on

 

 

 

 

that the sensor is visible.

 

 

 

 

 

 

 

the side of the new sensor. Press the new

 

 

 

 

 

 

 

 

3.

Plug the sensor harness into

the

transmitter

 

 

sensor’s pins into the three sockets.

 

 

 

 

7.

Carefully

replace the

cover

on

the sensor

 

assembly and

check

that no

components

are

 

 

touching the enclosure. Restore power and allow

 

 

assembly including the spacer washer.

 

 

 

 

 

 

 

 

 

 

a minimum of 30 minutes for stabilization before

 

8.

Install the sensor assembly into the enclosure and

 

 

 

adjustment.

 

 

 

 

 

 

 

 

 

 

 

tighten firmly.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4. Connect a DVM across TP10 and GT1 on the

 

9.

Reconnect the sensor harness to the transmitter,

 

 

transmitter assembly. Figure 7-12

 

 

 

 

 

 

install the transmitter into the housing and restore

 

 

 

 

 

 

 

5.

Locate

the

gain potentiometer

which

is on

the

 

 

power.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

sensor electronics directly behind one of the holes

 

10.

Allow one hour for the sensor to stabilize prior to

 

 

 

in the

sensor

body.

The

potentiometer

is

 

 

recalibration.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

accessible by inserting a jewelers screwdriver

 

11.

Recalibrate the sensor module.

 

 

 

 

 

 

 

 

 

 

 

through the hole in the sensor.

 

 

 

 

 

 

7.9.3.3

SENSOR OUTPUT ADJUSTMENT

 

 

 

6. Determine the correct value to be read at TP10

 

 

 

 

 

based on the following formula:

 

 

 

 

 

 

The following procedure is to be used when, during

VTP10 = 2(C/R) , where C = concentration of span gas,

 

normal calibration of a new sensor, the span voltage

 

and R = range of detection (100 PPM). The value

 

cannot be adjusted to a high enough level.

 

 

 

 

is negative for all models except 5100-06, and

 

 

 

 

 

 

 

 

 

 

5100-19.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SERVICE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page:

79

 

 

 

 

 

 

 

 

 

 

 

 

Page 12
Image 12
Sierra Monitor Corporation 5000 instruction manual Toxic Gas Transmitter Component Locator