English

2.3 Zeroing

Press the zeroing key to correct minor deviations from the zero point, e.g. caused by soiling on the scale.

Zero range approvable and non-approvable: -1% to +3% of the weighing range. If zeroing is not possible, the error message "Above zero range limit" or "Below zero range limit" appears for two seconds in the display.

Zero tracking

Zero tracking automatically corrects minor deviations in the zero display.

2.4 X10 display

Shows the weight value with an additional decimal point at ten times the resolution. Switch to the x10 display by pressing the Print key and then the Zeroing key, and hold them both down simultaneously.

Approvable scale

If the scale is approvable, the ten times resolution appears and switches off after 5 seconds.

Non-approved scale

If the scale is calibrated as non-approved, the ten times resolution appears constantly in the display. Press the key once to activate the ten times resolution function. Press the key again to deactivate the ten times resolution function.