CALIBRATION AND REPAIR
OFFSET ADJUSTMENT
The SR560's
CALIBRATION
There are four pots, which are used to calibrate the instrument. The pots adjust the
These pots should be adjusted to optimize the CMRR or null the offset when the front- end FET is replaced. Two of the pots adjust the CMRR and offset when the
First, the front panel offset pot must be set to zero:
•Adjust front panel Offset pot to read 0 VDC on pin 14, U407.
Next adjust the offset and CMRR for the case where the
•Couple = GND, Gain = 5 k, LOW NOISE: adjust P103 to null DC and output.
Now use a function generator as the source of a common mode signal:
•Apply 1 kHz 1 Vpp sine to both the A and B inputs.
•Couple = DC, source = A - B: adjust P102 to null sine wave output.
Now adjust the offset and CMRR for the case where the
•Couple = GND (remove signal from A and B inputs), Gain = 5 k, HIGH DR: adjust P104 to null DC and output.
•Apply 1 kHz 1 Vpp sine to both the A and B inputs.
•Couple = DC, source = A - B: adjust P101 to null sine wave output.
NOTE: In the above procedures, the gain of the
FRONT END REPLACEMENT
The most commonly damaged component is the
When replacing the FET, be certain that all eight pins are inserted into the socket, and observe the orientation of pin #1. After replacement adjust the CMRR and offset per the calibration procedure. More severely damaged
17