SRS Labs SR850 manual LIA Status Byte, Input/Resrv

Models: SR850

1 290
Download 290 pages 36.53 Kb
Page 208
Image 208

Remote Programming

LIA STATUS BYTE

bit

name

usage

 

0

INPUT/RESRV

Set when an INPUT or RESRV overload is

 

 

 

detected.

 

1

FILTR

Set when a FILTR overload is detected.

 

2

OUTPT

Set when an OUTPT overload is detected.

 

3

UNLK

Set when a reference unlock is detected.

 

4

RANGE

Set when the detection frequency switches

 

 

 

ranges (harmonic x ref. frequency decreases

 

 

 

below 199.21 Hz or increases above

 

 

 

203.12 Hz). Time constants above 30 s and

 

 

 

Synchronous filtering are turned off in the upper

 

 

 

frequency range.

 

5

TC

Set when the time constant is changed indirect-

 

 

 

ly, either by changing frequency range, dynamic

 

 

 

reserve or filter slope.

 

6

TRIG

Set when unit is triggered. Only if samples or

 

 

 

scans are triggered.

 

7

PLOT

Set when a plot is completed.

The LIA Status bits stay set until cleared by reading or by the CLS command.

ERROR STATUS BYTE

bit

name

usage

 

0

Prn/Plt Error

Set when an error occurs during printing or

 

 

 

plotting.

 

1

Backup Error

Set at power up when the battery backup has

 

 

 

failed.

 

2

RAM Error

Set when the RAM Memory test finds an error.

 

3

Disk Error

Set when an error occurs during a disk or file

 

 

 

operation.

 

4

ROM Error

Set when the ROM Memory test finds an error.

 

5

GPIB Error

Set when GPIB fast data transfer mode aborted.

 

6

DSP Error

Set when the DSP test finds an error.

 

7

Math Error

Set when an internal math error occurs.

The Error Status bits stay set until cleared by reading or by the CLS command.