Remote Programming

 

Time constants greater than 30s may NOT be set if the

 

harmonic x ref. frequency (detection frequency) exceeds 200 Hz. Time

 

constants shorter than the minimum time constant (based upon the filter

 

slope and dynamic reserve) will set the time constant to the minimum

 

allowed time constant. See the GAIN/TC menu section.

OFSL (?) {i}

The OFSL command sets or queries the low pass filter slope. The

 

parameter i selects 6 dB/oct (i=0), 12 dB/oct (i=1), 18 dB/oct (i=2) or

 

24 dB/oct (i=3).

SYNC (?) {i}

The SYNC command sets or queries the synchronous filter status. The

 

parameter i selects Off (i=0) or synchronous filtering below 200 Hz (i=1).

 

Synchronous filtering is turned on only if the detection frequency (refer-

 

ence x harmonic number) is less than 200 Hz.