Sunrise Medical 931157 Battery Wire Harness, If polarity is reversed correct battery wiring

Models: 931157

1 25
Download 25 pages 50.79 Kb
Page 9
Image 9
1.6 Battery Wire Harness

1.6 Battery Wire Harness

Check that the battery wire harness has the polarity cor- rect. Set the meter to dc volts and measure the connector with the red lead on the red wire contact (top of the con- nector) and the black lead on the black wire contact (bot- tom of the connector) as shown in (figure A1.6.1). If both battery wire harnesses have full voltage and correct polari- ty, then proceed to step 1.8. If voltage is absent proceed to step 1.7. If polarity is reversed correct battery wiring.

A1.6.1

If polarity is reversed correct battery wiring.1.7 Battery FuseIf polarity is reversed correct battery wiring.

 

1.7 Battery Fuse

 

 

 

 

 

 

Check that the battery fuse is in good condition. With the

 

 

 

 

 

batteries disconnected remove the fuse cap, inspect the

 

 

 

 

 

fuse to see if the fuse is blown. To make sure the fuse is

 

 

 

 

 

not blown, set the meter to ohms and measure the resist-

 

 

 

 

 

ance across the fuse. see (figure A1.7.1). If the meter

 

 

 

 

 

reads more than one ohm, change the fuse, or else pro-

 

 

 

 

 

ceed to the next step.

 

 

 

 

 

 

 

A1.7.1

 

 

 

 

 

 

 

 

 

 

If the meter reads more than one ohm, change the Battery fuse.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.8 Circuit Breaker Test

 

 

 

 

 

 

 

 

 

 

 

To check the circuit breaker set the meter to ohms and

 

 

 

 

 

measure the resistance across the circuit breaker as shown

 

 

 

 

 

in (figure A1.8.1). If the meter reads more than 1 ohm,

 

 

 

 

 

then change the circuit breaker, otherwise proceed to next

 

 

 

 

 

step.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A1.8.1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If the meter reads more than 1 ohm, then change the circuit breaker.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.9 Main Harness

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If the above steps did not correct the problem, change the

 

 

 

main harness.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If the previous steps did not correct the problem, change the main harness.

 

 

SUNRISE MEDICAL

ASPIRE SERVICE MANUAL

2003

PAGE 3

Page 9
Image 9
Sunrise Medical 931157 Battery Wire Harness, If polarity is reversed correct battery wiring, Battery Fuse, Main Harness