Tamron MST2000 Ex Tps Tpscal, Aprsrc Aprcon Ex Apz, Apzcal Ex Aps, Parameter, Description

Models: MST2000

1 33
Download 33 pages 61.13 Kb
Page 14
Image 14

OPERATION MANUAL

PROGRAM PARAMETERS

 

 

 

Parameter

Description

FactoryDefault

 

 

 

EX_TPS

TPSCAL

External Temperature Span. The External Temperature Span is the upper range setting for the external 4-20 milliamp ‘temperature’ input channel. The avail able range is:

English Unit Version: 0.000 to 990.0000 degrees F.

Metric Unit Ver sion: 0.000 to 990.0000 de grees C.

Temperature Span Calibration. The Temperature Span Calibration Value is the 20 milliamp (span) count value from the external 4-20 milliamp ‘temperature’ input channel. This parameter is used to calibrate the span point of the incoming 4-20 signal from the external temperature device. To calibrate the span of the incoming 4-20 signal, press the EDIT key. The screen will display IN20MA. Set the incoming signal for 20 milliamps (or connect an external 4-20 calibrator set for 20 milliamps) and press the EDIT key. The MST2000 will measure the external temperature input channel and save the new 20 milliamp reference to memory.

200.0000

21000

If an external temperature op tion is pre-installed and calibrated at the factory this will be custom set for that unit.

APRSRC

APRCON

EX_APZ

Absolute Pressure Source. Available assignments are 0 = CONstant 0 = CONstant

and 1 = EXTernal. þNote: MST2000 DP sen sor max i mum op er at ing static pres sure is 25 PSID.

Absolute Pressure Constant. The absolute pressure constant can

14.6960 for English.

be used for mass flow calculations when no external absolute pressure

1.01325 for Metric.

input is available and is entered as the average absolute pressure

 

inside the duct. The avail able range is:

 

English Unit Version: 0.000 to 100.0000 PSIA.

 

Metric Unit Ver sion: 0.000 to 100.0000 Bar.

 

External Absolute Pressure Zero. The External Absolute Pressure

0.000

Zero is the lower range (4 milliamp zero) setting for the external 4-20

 

milliamp ‘absolute pressure’ input channel. The avail able range is:

 

English Unit Version: 0.000 to 100.0000 PSIA.

þNote: MST2000 DP sensormaximum

Metric Unit Ver sion: 0.000 to 100.0000 Bar.

op er at ing static pres sure is 25 PSID.

APZCAL

EX_APS

Absolute Pressure Zero Calibration The Absolute Pressure Zero Calibration value is the 4 milliamp (zero) count value from the external 4-20 milliamp ‘absolute pressure’ input channel. This parameter is used to calibrate the zero point of the incoming 4-20 signal from the external 4-20 absolute pressure device. To calibrate the zero point of the incoming 4-20 signal, press the EDIT key. The screen will display IN4MA. Set the incoming signal for 4 milliamps (or connect an external 4-20 calibrator set for 4 milliamps) and press the EDIT key. The MST2000 will measure the external absolute pressure channel and save the new 4 milliamp reference to memory.

External Absolute Pressure Span. The External Absolute Pressure Span is the upper range setting for the external 4-20 milliamp ‘absolute pressure’ input channel. The available range is:

4175

If an external absolute pressure option is pre-installed and calibrated at the factory this will be custom set for that unit.

25.0000 for English.

1.7237 for Metric

English Unit Version: 0.000 to 100.0000 PSIA.

þNote: MST2000 DP sensormaximum

Metric Unit Ver sion: 0.000 to 100.0000 Bar.

op er at ing static pres sure is 25 PSID.

Brandt Instruments, Inc.

MST2000 Multivariable SMARTFLOW ® Transmitter

Page 15

Page 14
Image 14
Tamron MST2000 operation manual Ex Tps Tpscal, Aprsrc Aprcon Ex Apz, Apzcal Ex Aps, Parameter, Description, FactoryDefault