OPERATION MANUAL | TEST JACK |
TEST JACK
A Test Jack is stan dard on the Non I.S. Ap proved MST2000’s. It al lows the user to mon i tor the mA out- put of the unit with out dis con nect ing the loop. Re view di men sional draw ings on page 4. The Test Jack
is removed for all I.S. and Division 2 approved MST2000’s.
To monitor the mA output of the MST2000 you will need a precision milliammeter. Push the positive l ead from the meter into the positive jack (RED) and the negative lead from the meter into the negative jack (BLACK).
12. OPTIONAL MODULE INSTALLATION
The MST2000 Loop Powered Multivariable SMARTFLOW® Transmitter’s design is such that op- tional in put, dig i tal I/O and HART® com mu ni ca tion mod ules can be in stalled in the field. This al lows the user the ver sa til ity to change and adapt to dif fer ent ap pli ca tion needs. This sec tion contains guide lines for in stall ing the op tional mod ules and up dat ing the MST2000’s pro gram ming if nec es sary.Please re- view this section before attempting to make any upgrades. If there are any questions or problems please call the factory for assistance.
þNOTE:
&Any upgrades should be made in a clean and dust free environment.
&
12.1Isolated 4-20mA Input Mod ules (ISO Mod ule)
A.Thermo Brandt Part Number
a.The Iso lated
b.All iso lated
c.The Mod ule will be shipped in a sealed bag along with stand offs (2) and in struc tion sheet.
B.Installing the ISO Module Option.
a.Dis con nect all power from the MST2000.
b.De cide which slot the ISO Mod ule will fill. Re move the Op tion Cover La bel cov er ing that slot.
c.Take off the MST2000 Cover by re mov ing the 4 flat head screws. It is not nec es sary to re move the Main Board from the hous ing.
d.Lo cate the slot, con nec tor and two mount ing holes for the ISO Mod ule on the Main Board.
e.Re move the ISO Mod ule and two (2) stand offs from the bag.
f.Snap the Stand offs into the two holes lo cated on the Main Board as shown in the draw ing. Note the
ori en ta tion of the stand off.
g.
Brandt Instruments, Inc. | MST2000 Multivariable SMARTFLOW ® Transmitter | Page 23 |