Measuring Waveforms

TDS 500C, TDS 600B, & TDS 700C User Manual 3–141

Table 3–12: Standard masks

SONET
SDH ITU-T T1.102 Fibre channel Video Miscellaneous
None None None None None None
OC1/STM0
51.84 Mb/s DS–0 Sgl
Single 64.4 kb/s DS1
1.544 Mb/s FC133
Optical 132.8 Mb/s 4fsc NTSC
“D2” 143.18 Mb/s FDDI Halt 125 Mb/s
OC3/STM1
155.52 Mb/s DS-0 Dbl
Double 64 kb/s DS1A
2.048 Mb/s FC266
Optical 265.6 Mb/s 4:2:2
“D1” 270 Mb/s
OC12/STM4
622.08 Mb/s DS-0 Data
Data Contra 64 kb/s DS1C
3.152 Mb/s FC531
Optical 531.2 Mb/s
DS-0 Tmg
Timing 64 kb/s DS2
6.312 Mb/s FC1063
Optical 1.0625 Gb/s
E1 Sym
Sym Pair 2.048 Mb/s DS3
44.736 Mb/s FC133E
Electrical 132.7 Mb/s
E1 Coax
Coax Pair 2.048 Mb/s DS4NA
139.26 Mb/s FC266E
Electrical 265.6 Mb/s
E2
8.448 Mb/s DS4NA Mx
Max Output 139.26 Mb/s FC531E
Electrical 531.2 Mb/s
E3
34.368 Mb/s STS-1 Eye
Eye 51.84 Mb/s FC1063E
Electrical 1.0625 Gb/s
E4 0
Binary 0 139.26 Mb/s STS-1
Pulse 51.84 Mb/s
E4 1
Binary 1 139.26 Mb/s STS-3
155.52 Mb/s
E5 CEPT 565 Mb/s STS-3 Mx
Max Output 155.52 Mb/s
STM1E 0
Binary 0 155.52 Mb/s
STM1E 1
Binary 1 155.52 Mb/s
Optimizing Measurement Accuracy: SPC and Probe Cal

The TDS Oscilloscope provides three features that optimize measurement

accuracy. Signal Path Compensation (SPC) lets you compensate the internal

signal path used to acquire the waveforms and measure based on the ambient

temperature. Channel/Probe Deskew lets you compensate for the fact that signals

may come in from cables of different length. Probe Cal lets you compensate the

entire signal path, from probe tip to digitized signal, to improve the gain and

offset accuracy of the probe. This section tells you how to use both features.