3.0SPECIFICATIONS, WARRANTY

3.1Specifications

Table 2. Specifications for the converter and the Bypass Pump

Converter

Remotely mounted molybdenum converter with temperature

 

controller in a NEMA – 4 enclosure.

Converter Temperature

315oC 7oC with read-out on front panel of bypass pump

 

chassis.

Tube/Electrical Cable

Up to 50 feet (15 m) maximum length

Assembly

 

Power, Bypass Pump

120V~ 60 Hz, 230V~ 50 Hz 360 watts

Assembly

 

Power, Converter

120V~ 50/60 Hz, 60 watts

Assembly

 

Converter Efficiency

>96% for NO2

Weight, Bypass Pump

35 lbs (16 kg)

Assembly

 

Weight, Remote

30 lbs (11 kg)

Converter

 

Bypass Pump Chassis

9" x17" x23.6" (23cm x 43cm x 61cm)

Dimensions HxWxD

 

Remote Converter Assy

12"x19"x7" (31cm x 48cm x 18cm)

Dimensions HxWxD

 

Environmental

Installation Category (Over-voltage Category) II

 

Pollution Degree 2

Bypass Flowrate

800 cc/min 10%, each channel

Cal gas Flowrate

4000 cc/min (approx.)

Analog output

Converter temp analog output. 3.15 VDC = 315oC

05386 Rev. D

11

Page 15
Image 15
Teledyne 05386D manual SPECIFICATIONS, Warranty, Specifications