Model 200EU-NOy Option Manual Addendum
34 05386 Rev. D
SECONDARY MENU (The fixed characteristics of the system).
PRESS DISPLAY ACTION
SEL hold until p-n1 SET TO 0 – REVERSE ACTION
SEL hold until p-n2 SET TO 2 = TYPE J THERMOCOUPLE
SEL p-dF DIGITAL FILTER SET TO 5
SEL P-SL LOWER LIMIT SET TO -40 (-40ºC)
SEL P-SU UPPER LIMIT SET TO 350 (350ºC)
SEL P-AH ALARM SET TO 0
SEL P-An HYTERESIS SET TO 5
SEL P-dP PROG DECIMAL SET TO 0
SEL PVOF PROCESS OFFSET LEAVE AT 0
SEL SVOF SET POINT OFFS LEAVE AT 0
SEL P-F SET TO 0 (0 = oC, 1 = oF)
SEL P-n1 ALREADY CONFIGURED
SEL hold until RETURNS TO MAIN MENU
SEL PV/SV SET DISPLAY TO 315 C
SEL hold until P PROP BAND ALREADY CONFIGURED
SEL SCROLL TO A7 SET TO 1 (AUTOTUNE )