Model 200EU-NOy Option Manual Addendum

NOTE

The M200EU with the NOy Option does NOT have equivalency approval, and may not be used for EPA monitoring.

5.1 Calibration or Cal Check Procedure

The calibration of the instrument can be checked or adjusted using gas introduced through the calibration port “CAL IN” on the Bypass Pump Module. The calibration gas is routed to a TEE near the sample inlet port on the remote converter assembly.

Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero value. Table 3 details the zero calibration procedure.

Table 3 Manual Zero Calibration Procedure - Zero Gas Through Sample Port

Step

Action

Comment

Number

 

 

 

 

 

1.

Press CAL

The M200EU enters the calibrate mode.

 

 

 

2.

Wait For Stability

Wait for reading to stabilize at zero value.

 

Typ STB < 0.1 PPB

 

 

 

 

3.

Press ZERO

If you change your mind after pressing ZERO, you can

 

 

still press EXIT here without zeroing the instrument.

4.

Press ENTR

Pressing ENTR actually changes the calculation

 

 

equations.

5.

Press EXIT

M200EU returns to sampling. Immediately after

 

 

calibration, data is not added to the DAS averages.

 

 

 

16

05386 Rev. D

Page 20
Image 20
Teledyne 05386D manual Calibration or Cal Check Procedure