Model 200EU-NOy Option Manual Addendum
05386 Rev. D 33
Indications that would require re-programming are:
Failure / replacement of controller.
Replacement of thermocouple element.
Controller’s loss of PID parameters leading to instability of the temp setpoint.
In the event that the control parameters are changed or in the event that a new controller is
installed, it must be reprogrammed to suit the thermal characteristics of the instrument. The
controller includes an Autotuning feature, which can be used to properly set the PID factors.
To start the Autotune procedure:
1. Press Sel until A7 (AT) is displayed
2. Press Data until value = 1 (enable Autotune)
While Autotuning, the decimal point LED will blink.
NOTE: During the procedure, it is normal for the temp to overshoot up to 100°C.
When the decimal point LED turns off, Autotune is complete. No further adjustment
should be necessary.
Table 7 Temperature Controller Programming Guide
PRESS DISPLAY ACTION
SEL
DATA P PROP BAND UP/DOWN SET TO 5.0 (ENTER)
SEL
DATA i INTEGRAL SET TO 50
SEL
DATA d DERIVATIVE TIME SET TO 15
SEL
DATA AL LOW ALARM SETP SET TO 300
SEL
DATA AH HIGH ALARM SETP SET TO 330
SEL
DATA 7C CYCLE TIME SET TO 2 (FOR FAST SYSTEM)
SEL HYS HYSTERESIS SET TO 0.3
SEL LOC LOCK 0 = OPEN 1 = LOCKED
2 = SV ONLY OPEN
NOTE:
DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC.