6.0MAINTENANCE

6.1Maintenance Schedule

The schedule for preventative maintenance is presented in Table 6 below.

Table 6 Preventative Maintenance Calendar

Item

Maintenance Interval

 

 

Zero/Span Calibration

Annually or after maintenance or repairs

 

 

Zero/Span Checks

Daily

 

 

Particulate Filter

Weekly as needed

 

 

Ozone Flow

Check every year and replace o-ring and

 

orifice in reaction cell

Ozone Dryer Filter-DFU

Replace every year

 

 

Sample Flow

Check every 2 months

 

 

Converter

Check efficiency every 2 months

 

 

Pneumatic Lines

Check every 3 months

 

 

Factory Calibration

Calibrate each year or after repairs

 

 

Leak Check

Check every 6 months

 

 

Replace bypass flow orifices

Every year / as needed

 

 

Rebuild bypass pump

Every year

 

 

05386 Rev. D

19