Teledyne 2002 instruction manual Power Entry Module, Model Rear Panel Detail

Models: 2002

1 54
Download 54 pages 12.52 Kb
Page 20
Image 20
4.4 Power Entry Module

4.4 Power Entry Module

The AC power for the Model 2002 enters the instrument through a power entry module that contains a fuse, on/off switch and an IEC 320 power inlet.

The fuse is rated for 250 V, 1/4 A. It can be accessed by unplugging the AC cord and prying the fuse compartment open with the tab and slot in the power inlet chamber. There is a spare fuse in the compartment within the fuse holder.

LOW SETPOINT

OPTION BOARD

HIGH SETPOINT

RELAY

CONNECTOR (MALE)

RELAY

TRANSDUCER CONN.

STANDARD

(FEMALE)

I/O CONNECTOR

Analog 0-10V OPTION

BOARD CONNECTOR

Pin# Connection

1Analog Out 1 (1024 Torr) High

2Analog Out 1 (1024 Torr) Shield

3Analog Out 1 (1024 Torr) Low

7Analog Out 2 (1000 mTorr) Low

8Analog Out 2 (1000 mTorr) High

9Analog Out 2 (1000 mTorr) Shield

4-20mA OPTION BOARD CONNECTOR

Pin#

Connection

 

1

I-Loop Out (+) 1024 Torr

3

I-Loop In

(-) 1024 Torr

8

I-Loop Out (+) 1000 mTorr

7

I-Loop In

(-) 1000 mTorr

Model 2002

Rear Panel Detail

Figure 4.2

SENSOR CONNECTOR

Pin# Connection

1Sensor

2Bridge (sense)

3Bridge-Power

4Reference

5Common

6Common (sense)

7N/C

8N/C

I/O CONNECTOR

Pin# Connection

1Analog Output

2Analog Common

3High Setpoint

4Low Setpoint

5Digital Common

6Remote Zero

7+5V

page 20 Model 2002

Page 20
Image 20
Teledyne 2002 instruction manual Power Entry Module, Model Rear Panel Detail