4.4 Power Entry Module
The AC power for the Model 2002 enters the instrument through a power entry module that contains a fuse, on/off switch and an IEC 320 power inlet.
The fuse is rated for 250 V, 1/4 A. It can be accessed by unplugging the AC cord and prying the fuse compartment open with the tab and slot in the power inlet chamber. There is a spare fuse in the compartment within the fuse holder.
LOW SETPOINT | OPTION BOARD | HIGH SETPOINT |
RELAY | CONNECTOR (MALE) | RELAY |
TRANSDUCER CONN. | STANDARD |
(FEMALE) | I/O CONNECTOR |
Analog
BOARD CONNECTOR
Pin# Connection
1Analog Out 1 (1024 Torr) High
2Analog Out 1 (1024 Torr) Shield
3Analog Out 1 (1024 Torr) Low
7Analog Out 2 (1000 mTorr) Low
8Analog Out 2 (1000 mTorr) High
9Analog Out 2 (1000 mTorr) Shield
Pin# | Connection |
|
1 | ||
3 | ||
8 | ||
7 |
Model 2002
Rear Panel Detail
Figure 4.2
SENSOR CONNECTOR
Pin# Connection
1Sensor
2Bridge (sense)
3
4Reference
5Common
6Common (sense)
7N/C
8N/C
I/O CONNECTOR
Pin# Connection
1Analog Output
2Analog Common
3High Setpoint
4Low Setpoint
5Digital Common
6Remote Zero
7+5V
page 20 Model 2002