Communications Option Board

SECTION 6

Communications Option Board

Both RS-232 and RS-485 communication interface boards are available as an option for the Model 2002.The communication option boards allow data to be output to a host computer with the appropiate interface. Connection to the RS-232 or RS-485 communication interface boards are via a 9 pin D-subminiature female connector.

6.1RS-232-E Interface Specifications

Format

EIA standard RS-232-E, full-duplex, no handshaking, asynchronous

Data Rate

9600 baud

Character Length

Eight data bits

Parity

No parity

Stop bits

One stop bit

A common application of the RS232 version of the HPM-2002, is to connect the pressure gauge directly to the serial port of a PC. This is done by first wiring a communication cable in the

Figure 6.1

manner shown below.

PIN

MNEMONIC

SIGNAL

DESCRIPTION

 

 

 

 

 

 

 

 

2

TX

Transmit Data

Transmits data within RS-232-E voltage levels

 

 

3

RX

Receive Data

Accepts data within RS-232-E voltage levels.

 

 

5

SG

Signal Ground

Establishes the common ground reference

 

 

 

 

 

potential for all interchange circuits

 

 

when

 

 

communicating via RS-

 

 

232-E.

 

 

 

 

 

4

CTS

Clear to send

Host computer ready to accept data. (Reserved

 

 

 

 

 

for future use, not implemented by present

 

 

 

 

 

software.

 

 

8

RDY

Ready

Ready to receive data (always high, not

 

 

 

 

 

implemented by present software.

 

 

 

 

 

 

 

 

 