Device Status

When requested to transmit its status the Model 2002 responds with a five digit number which is explained in the following:

digit#:

1

2

3

4

5

Serial Receiver Overflow

= 4

 

 

 

 

Main Board EEPROM Error

= 2

 

 

 

 

Probe EEPROM (any) Error

= 1

 

 

 

 

Probe EEPROM Not Responding

 

= 8

 

 

 

Probe EEPROM Read Error

 

= 1

 

 

 

Probe EEPROM Checksum Error

 

 

= 4

 

 

Probe EEPROM Verification Error

 

 

= 2

 

 

Probe EEPROM Identification Error

 

 

= 1

 

 

Communications Syntax Error

 

 

 

= 4

 

Piezo Sensor Bad (voltage out of range)

 

 

 

= 2

 

Pirani Sensor Bad (voltage out of range)

 

 

 

= 1

 

High Setpoint Alarm (pressure exceeds setpoint)

 

 

 

= 4

Low Setpoint Alarm (pressure less than setpoint)

 

 

= 2

Gas# Changed (not the same as when unit last calibrated)

 

= 1

page 34 Model 2002