Device Status
When requested to transmit its status the Model 2002 responds with a five digit number which is explained in the following:
digit#: | 1 | 2 | 3 | 4 | 5 |
Serial Receiver Overflow | = 4 |
|
|
|
|
Main Board EEPROM Error | = 2 |
|
|
|
|
Probe EEPROM (any) Error | = 1 |
|
|
|
|
Probe EEPROM Not Responding |
| = 8 |
|
|
|
Probe EEPROM Read Error |
| = 1 |
|
|
|
Probe EEPROM Checksum Error |
|
| = 4 |
|
|
Probe EEPROM Verification Error |
|
| = 2 |
|
|
Probe EEPROM Identification Error |
|
| = 1 |
|
|
Communications Syntax Error |
|
|
| = 4 |
|
Piezo Sensor Bad (voltage out of range) |
|
|
| = 2 |
|
Pirani Sensor Bad (voltage out of range) |
|
|
| = 1 |
|
High Setpoint Alarm (pressure exceeds setpoint) |
|
|
| = 4 | |
Low Setpoint Alarm (pressure less than setpoint) |
|
| = 2 | ||
Gas# Changed (not the same as when unit last calibrated) |
| = 1 |
page 34 Model 2002